Insira pelo menos 2 caracteres para iniciar a busca.
  • Histórico

TP analyzer
Liquiline System CA80TP

Liquiline System CA80TP - TP analyzer for environmental monitoring ©Endress+Hauser
Inside view of the TP analyzer with cooling module ©Endress+Hauser

The total phosphorus concentration of a liquid can be determined using chemical and thermal digestion and the colorimetric measuring principle. This video shows how this measuring principle works.

Total phosphorus analyzer for environmental monitoring, industrial and municipal wastewater

Phosphorus is the decisive factor for excessive algae and plant growth in water bodies, that's why continuous monitoring of this parameter is essential. The Liquiline System TP analyzer offers precise online measurement of total phosphorus and ensures highest process safety. Automatic calibration and cleaning save you operating costs while advanced diagnostics with remote access support you in providing process documentation to the authorities.

  • Produtos simples

  • Fácil de escolher, instalar e operar

Excelência técnica

Simplicidade

  • Produtos padrão

  • Confiável, robusto e baixa manutenção

Excelência técnica

Simplicidade

  • Produtos com tecnologia de ponta

  • Altamente funcional e prático

Excelência técnica

Simplicidade

  • Produtos especializados

  • Desenvolvidos para aplicações exigentes

Excelência técnica

Simplicidade

Variável

Versões FLEX Excelência técnica Simplicidade
  • F
  • L
  • E
  • X

Versão Fundamental

Atende suas necessidades básicas de medição

Excelência técnica
Simplicidade
  • F
  • L
  • E
  • X

Versão Lean

Lida com seus principais processos facilmente

Excelência técnica
Simplicidade
  • F
  • L
  • E
  • X

Seleção Extended

Otimiza seus processos com tecnologias inovadoras

Excelência técnica
Simplicidade
  • F
  • L
  • E
  • X

Versão Xpert

Controla suas aplicações mais difíceis

Excelência técnica
Simplicidade

Variável

  • Especificações resumidas

    • Measurement range

      0.05 to 10 mg/l
      0.5 to 50 mg/l

    • Process temperature

      4 to 40 °C (39.2 to 104 °F)

    • Process pressure

      At atmospheric pressure, < 0.2 bar

please wait Loading alternative products

Seleção rápida

Escolha uma das opções sugeridas
Preço de venda
  • <%label%>
  • <%feature%>
  • Process Connection
  • <%message%>

ou

Configuração individual

Continue sua seleção entre todas as opções possíveis
  • Benefícios

    • Molybdenum blue method according to EN ISO 6878:2004 guarantees direct comparability to laboratory results

    • Meeting industry needs: The analyzer is perfectly suited for environmental monitoring, industrial and municipal wastewater

    • Fast and easy process...

  • Campo de aplicação

    The Liquiline System TP analyzer monitors:

    • Inlet and outlet of wastewater treatment plants for documentation purposes and calculation of the cleaning capacity

    • Discharges of industrial wastewater treatment plants to determine discharge fees and to...

Características e especificações

Voltar

Details

  • Measuring principle

    Colorimetric

  • Characteristic

    Analyzer for total phosphorus

  • Size

    Housing (open version):

    793 x 530 x 417 mm

    31.22 x 20.87 x 16.42 in


    Housing (closed version):

    793 x 530 x 463 mm

    31.22 x 20.87 x 18.23 in


    Housing with base:

    1723 x 530 x 463 mm

    67.83 x 20.87 x 18.23 in

  • Design

    Open design, cabinet and stand housing
    High-Performance plastic ASA-PC, additional stand coated steel

  • Process temperature

    4 to 40 °C (39.2 to 104 °F)

  • Ambient temperature

    5 to 40 °C (41 to 104 °F)

    Outdoor version: - 20 to 40 °C (-4 to 104 °F)

  • Process pressure

    At atmospheric pressure, < 0.2 bar

  • Sample flow rate

    Min. 30 ml/min

  • Consistency of the sample

    Low solid content (TS<50 mg/l), homogeneous

  • Specials

    Easy upgrade to measuring station with up to four digital Memosens sensors
    Automatic calibration and cleaning
    User-configurable measuring, cleaning and calibration intervals

    Self-priming analyzer with optical dosing unit l
    Dilution module (optional)
    Digital communication for remote access

  • Application

    Monitoring and optimization of the cleaning capacity of municipal and industrial wastewater treatment plants
    Monitoring of the wastewater treatment plant outlet for documentation purposes
    Monitoring and optimization of activated sludge basins

  • Power supply

    100 to 120 VAC / 200 to 240 VAC ± 10%
    24 VDC ± 10%
    50 ± 1 or 60 ± 1,2 Hz

  • Output / communication

    2x 0/4 to 20 mA
    Webserver, Modbus, Ethernet/IP, Profibus DP

  • Input

    1 or 2 measuring channel
    1 to 4 digitale sensor inputs for sensors with Memosens protocol (optional)

  • Measurement range

    0.05 to 10 mg/l
    0.5 to 50 mg/l

  • Consumables

    Reagents and standard solutions CY80TP are necessary for the operation