Especificações resumidas
-
Accuracy
class A acc. to IEC 60751
class B acc. to IEC 60751 -
Response time
depending on configuration
63% rt = 20 s -
Max. process pressure (static)
at 20 °C: 500 bar (7.252 psi)
depends on configuration -
Operating temperature range
PT100 WW:
-200 °C … 600 °C
(-328 °F … 1.112 °F)
PT100 TF:
-50 °C … 200 °C
(-58 °F … 392 °F) -
Max. immersion length on request
up to 22.5" (571 mm)
others on request
Campo de aplicação
The robust thermometer is designed for use in demanding and safety relevant applications e.g. in Chemical, Oil & Gas and Energy industry. Harsh environments, corrosive substances and highest pressures can be handled by the use of robust thermowells and special materials. A optional head transmitter with all common communication protocols makes the device ready to use with enhanced measurement accuracy and reliability compared to directly wired sensors. Flexible configuration possible.
The T14 temperature sensor is a RTD assembly installed in Thermowell and designed for use in harsh environments and explosive areas, due to his FM and CSA approvals.
The sensor is made up of a measurement probe with an insulated RTD element, sheath and a thermowell made of bar-stock material.
The sensor assembly can be used in process industries such as:
Chemicals
Petrochemical
Refineries
Offshore Platforms
Benefícios
FM/CSA XP Class I, Div. 1 approved temperature assemblies for maximum safety
One source shopping for temperature measurement solutions. World class transmitter with integrated sensor offering for heavy process industry applications
Remove and install straight out of the box!
Improved galvanic isolation on most devices (2 kV)
Simplified model structure: Competitively priced, offers great value. Easy to order and reorder. A single model number includes sensor, thermowell and transmitter assembly for a complete point solution
All iTEMP transmitters provide long term stability ≤0.05% per year
-
Produtos com tecnologia de ponta
- Altamente funcional e prático
Excelência técnica
Simplicidade
-
Produtos especializados
- Desenvolvidos para aplicações exigentes
Excelência técnica
Simplicidade
Variável
| Versões FLEX | Excelência técnica | Simplicidade |
|---|---|---|
Versão Fundamental Atende suas necessidades básicas de medição |
Excelência técnica
|
Simplicidade
|
Versão Lean Lida com seus principais processos facilmente |
Excelência técnica
|
Simplicidade
|
Seleção Extended Otimiza seus processos com tecnologias inovadoras |
Excelência técnica
|
Simplicidade
|
Versão Xpert Controla suas aplicações mais difíceis |
Excelência técnica
|
Simplicidade
Variável |
-
Measuring principle
Resistance Temperature Detector
-
Characteristic / Application
Explosion Proof US style
modular temperature assembly
for heavy duty applications
flanged process connection
with extension
incl. thermowell -
Thermowell / protection tube
bar stock (drilled)
-
Insert / probe
mineral insulated (MI), flexible
PTFE-insulated, rigid -
Outer diameter protection tube / Insert
7/8'' (22,23 mm)
17/16'' (26,99 mm) -
Max. immersion length on request
up to 22.5" (571 mm)
others on request -
Material protection tube/ thermowell
316/316L
others on request -
Optional coating
Available on request
-
Process connection
flange:
ASME 1" 150 RF (B16.5)
ASME 1" 300 RF (B16.5)
ASME 1" 600 RF (B16.5)
ASME 1" 900/1500 RF (B16.5)
ASME 1,5" 150 RF (B16.5)
ASME 1,5" 300 RF (B16.5)
ASME 1,5" 600 RF (B16.5)
ASME 1,5" 900/1500 RF (B16.5)
ASME 2" 150 RF (B16.5)
ASME 2" 300 RF (B16.5)
ASME 2" 600 RF (B16.5)
ASME 2" 900/1500 RF (B16.5) -
Tip shape
straight
tapered -
Surface roughness Ra
32 μin. (0.80 μm)
-
Operating temperature range
PT100 WW:
-200 °C … 600 °C
(-328 °F … 1.112 °F)
PT100 TF:
-50 °C … 200 °C
(-58 °F … 392 °F) -
Max. process pressure (static)
at 20 °C: 500 bar (7.252 psi)
depends on configuration -
Accuracy
class A acc. to IEC 60751
class B acc. to IEC 60751 -
Response time
depending on configuration
63% rt = 20 s -
Integration head transmitter
yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
-
Ex - approvals
FM XP DIP
FM XP NI DIP
CSA XP DIP
CSA XP NI DIP
FM/CSA XP DIP
FM/CSA XP NI DIP
CSA C/US XP, DIP
cCSAus General Purpose -
Certification
SIL (transmitter only)
-
-
Informações Técnicas (TI)
T13, T14 and T15
-
Informações Técnicas (TI)
Versão em inglês - 01/2023New version available in EnglishExplosion proof RTD assemblies in thermowells with spring loaded insert
and enclosure for process industryEN PT 10/01/202313/01/202131/05/201812/05/201526/10/2009Idioma:
Inglês
Versão:
10/01/2023
Tamanho do arquivo:
1.4 MB
Nome do arquivo:
TI00126REN_1623-00.pdf
-
-
-
Instruções curtas (KA)
T14 Compact Instructions
-
Instruções curtas (KA)
Versão em inglês - 12/2012New version available in EnglishExplosion proof RTD assembly in flanged thermowell
EN PT 01/12/201201/07/200701/03/2007Idioma:
Inglês
Versão:
01/12/2012
Tamanho do arquivo:
671.9 KB
Nome do arquivo:
KA00237rae_1312.pdf
-
-
-
Instruções de Instalação (EA)
iTEMP TMT72 Transmitter HART
-
Instruções de Instalação (EA)
Versão em inglês - 05/2026New version available in EnglishiTEMP TMT72 Transmitter HART
Bild für SFTIdioma:
Inglês
Versão:
06/05/2026
Tamanho do arquivo:
135.3 KB
Nome do arquivo:
Bild TMT72.pdf
-
-
-
Instruções de Instalação (EA)
iTEMP 85 FOUNDATION Fieldbus Protokoll
-
Instruções de Instalação (EA)
Versão em inglês - 05/2026New version available in EnglishiTEMP 85 FOUNDATION Fieldbus Protokoll
Bild für SFTIdioma:
Inglês
Versão:
06/05/2026
Tamanho do arquivo:
154 KB
Nome do arquivo:
Bild TMT85.pdf
-
-
-
Instruções de Instalação (EA)
EA Klemmenblock für TS211
-
Instruções de Instalação (EA)
Versão em inglês - 12/2025New version available in EnglishEA Klemmenblock für TS211
Idioma:
Inglês
Versão:
02/12/2025
Tamanho do arquivo:
322.1 KB
Nome do arquivo:
EA_TS211_Klemmblock _ 7166054x _en_we.pdf
-
-
-
Instruções de Instalação (EA)
iTEMP TMT31, Transmitter 4-20mA
-
Instruções de Instalação (EA)
Versão em inglês - 01/2021New version available in EnglishiTEMP TMT31, Transmitter 4-20mA
EN PT 01/01/2021Idioma:
Inglês
Versão:
01/01/2021
Tamanho do arquivo:
119.6 KB
Nome do arquivo:
Bild TMT31.pdf
-
-
-
Instruções de Instalação (EA)
iTEMP TMT82 Transmitter HART
-
Instruções de Instalação (EA)
Versão em inglês - 05/2026New version available in EnglishiTEMP TMT82 Transmitter HART
Bild für SFTIdioma:
Inglês
Versão:
06/05/2026
Tamanho do arquivo:
178.9 KB
Nome do arquivo:
Bild TMT82.pdf
-
-
-
Instruções de Instalação (EA)
iTEMP TMT84 PROFIBUS® PA
-
Instruções de Instalação (EA)
Versão em inglês - 05/2026New version available in EnglishiTEMP TMT84 PROFIBUS® PA
Bild für SFTIdioma:
Inglês
Versão:
06/05/2026
Tamanho do arquivo:
263.2 KB
Nome do arquivo:
Bild TMT84.pdf
-
-
-
Instruções de Segurança Ex (XA)
T13, T14, T53, T54
-
Instruções de Segurança Ex (XA)
Versão em inglês - 08/2019New version available in EnglishCSA:
Control Drawing XP, NI, DIP
160100118Idioma:
Inglês
Versão:
26/08/2019
Tamanho do arquivo:
65.5 KB
Nome do arquivo:
160100118.pdf
-
-
-
Instruções de Segurança Ex (XA)
Temperature assemblies T13, T14, T15, T53, T54, T55 Control drawing CSA (XP, DIP)
-
Instruções de Segurança Ex (XA)
Versão em inglês - 12/2022New version available in EnglishInstallation notes for CSA approved apparatus
Idioma:
Inglês
Versão:
19/12/2022
Tamanho do arquivo:
97 KB
Nome do arquivo:
XA02075TEN_03.22.pdf
-
Escolha uma das opções sugeridas
-
-- selected coding-- selected label99.999,00<%label%>
-
Procurando mais opções? Use o configurador individual para continuar.
-
-- selected coding-- selected label99.999,00<%feature%>
-
<%label%>99.999,00
-
Procurando mais opções? Use o configurador individual para continuar.
-
Process Connection
-
<%label%>99.999,00
-
<%label%>99.999,00
-
-
<%feature%><%coding%><%label%>
- Diagramas CAD Diagramas CAD
- Diagramas CAD
-
Preço por peça
-
<%range%><%price%>
-
- Acessórios opcionais
Configuração individual
Geralmente são fornecidos juntos
Acessórios
Produtos relacionados