Especificações resumidas
-
Max. measurement error
Volume flow (liquid): ±0.75 %
Volume flow (steam, gas): ±1.00 %
Mass flow (saturated steam): ±1.7% (temperature compensated); ±1.5% (temperature/pressure compensated)
Mass flow (superheated steam, gas): ±1.5 (temperature/pressure compensated); ±1.7% (temperature compensated + external pressure compensation)
Mass flow (liquid): ±0.85% -
Measuring range
Liquid: 0.2 to 1895 m³/h (0.15 to 1115 ft³/min)
depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)
Steam, gas: 1.5 to 25285 m³/h (0.9 to 14880 ft³/min)
depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a) -
Medium temperature range
Standard: –40 to +260 °C (–40 to +500 °F)
High/low temperature (option): –200 to +400 °C (–328 to +752 °F) -
Max. process pressure
PN 250, Class 1500, 40K
-
Wetted materials
Measuring tube: 1.4408 (CF3M)
DSC sensor: UNS N07718 similar to Alloy 718, 2.4668
Process connection: 1.4404/F316/F316L
Campo de aplicação
Prowirl O is ideally suited for reliable process control in demanding gas and steam applications with high process pressure. Moreover, its design ensures maximum safety in main and ancillary processes. With genuine loop-powered technology, Prowirl O 200 enables cost-effective and seamless integration into existing infrastructures. It offers highest operational safety in hazardous areas. Heartbeat Technology enables process safety at all times.
Preferred measuring principle for wet/saturated/superheated steam, gases & liquids (also cryogenic)
The specialist for applications with high process pressure
Device properties
Saturated steam mass flow up to PN 250 (Class 1500)
Full compliance with NACE (MR0175/MR0103)
Flexible positioning of pressure cell
Display module with data transfer function
Robust dual-compartment housing
Plant safety: worldwide approvals (SIL, Haz. area)
Benefícios
Better process control – integrated temperature and pressure measurement for steam and gases
Increased mechanical integrity for flow measurement – special sensor design
Same accuracy down to Re 10 000 – most linear Vortex meter body
Long-term stability – robust drift-free capacitive sensor
Convenient device wiring – separate connection compartment, various Ethernet options
Safe operation – no need to open the device
Integrated verification – Heartbeat Technology
Vídeos
-
Unboxing and commissioning of Proline Prowirl 200 vortex flowmeter with integrated temperature and pressure measurement
-
The Vortex Flow Measuring Principle
-
HistoROM data storage concept
-
This film shows that Prowirl 200 is the most robust vortex flowmeter worldwide even under extreme process conditions.
Seleção Extended
-
Produtos com tecnologia de ponta
- Altamente funcional e prático
Excelência técnica
Simplicidade
-
Produtos com tecnologia de ponta
- Altamente funcional e prático
Excelência técnica
Simplicidade
-
Produtos especializados
- Desenvolvidos para aplicações exigentes
Excelência técnica
Simplicidade
Variável
| Versões FLEX | Excelência técnica | Simplicidade |
|---|---|---|
Versão Fundamental Atende suas necessidades básicas de medição |
Excelência técnica
|
Simplicidade
|
Versão Lean Lida com seus principais processos facilmente |
Excelência técnica
|
Simplicidade
|
Seleção Extended Otimiza seus processos com tecnologias inovadoras |
Excelência técnica
|
Simplicidade
|
Versão Xpert Controla suas aplicações mais difíceis |
Excelência técnica
|
Simplicidade
Variável |
Predecessor
-
Measuring principle
Vortex
-
Product headline
Flowmeter optimized for requirements of high-pressure mating pipes.
Better process control – integrated temperature and pressure measurement for steam and gases.
The specialist for applications with high process pressure. -
Sensor features
Increased mechanical integrity for flow measurement – special sensor design. Same accuracy down to Re 10 000 – most linear Vortex meter body. Long-term stability – robust drift-free capacitive sensor.
Saturated steam mass flow up to PN 250 (Class 1500). Full compliance with NACE (MR0175/MR0103). Flexible positioning of pressure cell. -
Transmitter features
Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology.
Display module with data transfer function. Robust dual-compartment housing. Plant safety: worldwide approvals (SIL, Haz. area). -
Nominal diameter range
DN 15 to 300 (½ to 12")
-
Wetted materials
Measuring tube: 1.4408 (CF3M)
DSC sensor: UNS N07718 similar to Alloy 718, 2.4668
Process connection: 1.4404/F316/F316L -
Measured variables
Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature
-
Max. measurement error
Volume flow (liquid): ±0.75 %
Volume flow (steam, gas): ±1.00 %
Mass flow (saturated steam): ±1.7% (temperature compensated); ±1.5% (temperature/pressure compensated)
Mass flow (superheated steam, gas): ±1.5 (temperature/pressure compensated); ±1.7% (temperature compensated + external pressure compensation)
Mass flow (liquid): ±0.85% -
Measuring range
Liquid: 0.2 to 1895 m³/h (0.15 to 1115 ft³/min)
depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)
Steam, gas: 1.5 to 25285 m³/h (0.9 to 14880 ft³/min)
depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a) -
Max. process pressure
PN 250, Class 1500, 40K
-
Medium temperature range
Standard: –40 to +260 °C (–40 to +500 °F)
High/low temperature (option): –200 to +400 °C (–328 to +752 °F) -
Ambient temperature range
Compact version (standard): –40 to +80 °C (–40 to +176 °F)
Compact version (option): –50 to +80 °C (–58 to +176 °F)
Remote version (standard): –40 to +85 °C (–40 to +185 °F)
Remote version (option): –50 to +85 °C (–58 to +185 °F) -
Sensor housing material
Sensor connection housing: AlSi10Mg, coated; 1.4408 (CF3M)
-
Transmitter housing material
AlSi10Mg, coated; 1.4404 (316L)
-
Degree of protection
Compact version: IP66/67, type 4X enclosure
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, type 4X enclosure -
Display/Operation
4‐line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available -
Outputs
4‐20 mA HART (passive)
4‐20 mA (passive)
Pulse/frequency/switch output (passive) -
Inputs
4‐20 mA (passive)
-
Digital communication
HART, PROFIBUS PA, FOUNDATION Fieldbus, PROFINET over Ethernet-APL, Modbus TCP over Ethernet-APL
-
Power supply
DC 12 to 35 V (4‐20 mA HART with/without pulse/frequency/switch output)
DC 12 to 30 V (4‐20 mA HART, 4‐20 mA)
DC 12 to 35 V (4‐20 mA HART, pulse/frequency/switch output, 4‐20 mA input)
DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch output) -
Hazardous area approvals
ATEX, IECEx, cCSAus, JPN, KC
-
Product safety
CE, C-TICK, EAC, UK Ex
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation) -
Marine approvals and certificates
ABS, LR, BV, DNV, CCS
-
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
NACE MR0175/MR0103, PMI (on request); only Class 900/1500: welding test acc. to ISO 15614‐1, similar to ASME IX (on request)
-
Measuring principle
Vortex
-
Product headline
Flowmeter optimized for requirements of high-pressure mating pipes.
Better process control – integrated temperature and pressure measurement for steam and gases.
The specialist for applications with high process pressure. -
Sensor features
Increased mechanical integrity for flow measurement – special sensor design. Same accuracy down to Re 10 000 – most linear Vortex meter body. Long-term stability – robust drift-free capacitive sensor.
Saturated steam mass flow up to PN 250 (Class 1500). Full compliance with NACE (MR0175/MR0103). Flexible positioning of pressure cell. -
Transmitter features
Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology.
Display module with data transfer function. Robust dual-compartment housing. Plant safety: worldwide approvals (SIL, Haz. area). -
Nominal diameter range
DN 15 to 300 (½ to 12")
-
Wetted materials
Measuring tube: 1.4408 (CF3M)
DSC sensor: UNS N07718 similar to Alloy 718, 2.4668
Process connection: 1.4404/F316/F316L -
Measured variables
Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature
-
Max. measurement error
Volume flow (liquid): ±0.75 %
Volume flow (steam, gas): ±1.00 %
Mass flow (saturated steam): ±1.7% (temperature compensated); ±1.5% (temperature/pressure compensated)
Mass flow (superheated steam, gas): ±1.5 (temperature/pressure compensated); ±1.7% (temperature compensated + external pressure compensation)
Mass flow (liquid): ±0.85% -
Measuring range
Liquid: 0.2 to 1895 m³/h (0.15 to 1115 ft³/min)
depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)
Steam, gas: 1.5 to 25285 m³/h (0.9 to 14880 ft³/min)
depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a) -
Max. process pressure
PN 250, Class 1500, 40K
-
Medium temperature range
Standard: –40 to +260 °C (–40 to +500 °F)
High/low temperature (option): –200 to +400 °C (–328 to +752 °F) -
Ambient temperature range
Compact version (standard): –40 to +80 °C (–40 to +176 °F)
Compact version (option): –50 to +80 °C (–58 to +176 °F)
Remote version (standard): –40 to +85 °C (–40 to +185 °F)
Remote version (option): –50 to +85 °C (–58 to +185 °F) -
Sensor housing material
Sensor connection housing: AlSi10Mg, coated; 1.4408 (CF3M)
-
Transmitter housing material
AlSi10Mg, coated; 1.4404 (316L)
-
Degree of protection
Compact version: IP66/67, type 4X enclosure
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, type 4X enclosure -
Display/Operation
4‐line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available -
Outputs
4‐20 mA HART (passive)
4‐20 mA (passive)
Pulse/frequency/switch output (passive) -
Inputs
4‐20 mA (passive)
-
Digital communication
HART, PROFIBUS PA, FOUNDATION Fieldbus, PROFINET over Ethernet-APL, Modbus TCP over Ethernet-APL
-
Power supply
DC 12 to 35 V (4‐20 mA HART with/without pulse/frequency/switch output)
DC 12 to 30 V (4‐20 mA HART, 4‐20 mA)
DC 12 to 35 V (4‐20 mA HART, pulse/frequency/switch output, 4‐20 mA input)
DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch output) -
Hazardous area approvals
ATEX, IECEx, cCSAus, JPN, EAC, UK Ex, KC
-
Product safety
CE, C-TICK, EAC
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation) -
Marine approvals and certificates
ABS, LR, BV, DNV, CCS
-
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
NACE MR0175/MR0103, PMI (on request); only Class 900/1500: welding test acc. to ISO 15614‐1, similar to ASME IX (on request)
-
Measuring principle
Vortex
-
Product headline
Flowmeter optimized for requirements of high-pressure mating pipes.
Better process control – integrated temperature and pressure measurement for steam and gases.
The specialist for applications with high process pressure. -
Sensor features
Increased mechanical integrity for flow measurement – special sensor design. Same accuracy down to Re 10 000 – most linear Vortex meter body. Long-term stability – robust drift-free capacitive sensor.
Saturated steam mass flow up to PN 250 (Class 1500). Full compliance with NACE (MR0175/MR0103). Flexible positioning of pressure cell. -
Transmitter features
Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology.
Display module with data transfer function. Robust dual-compartment housing. Plant safety: worldwide approvals (SIL, Haz. area). -
Nominal diameter range
DN 15 to 300 (½ to 12")
-
Wetted materials
Measuring tube: 1.4408 (CF3M)
DSC sensor: UNS N07718 similar to Alloy 718, 2.4668
Process connection: 1.4404/F316/F316L -
Measured variables
Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature
-
Max. measurement error
Volume flow (liquid): ±0.75 %
Volume flow (steam, gas): ±1.00 %
Mass flow (saturated steam): ±1.7% (temperature compensated); ±1.5% (temperature/pressure compensated)
Mass flow (superheated steam, gas): ±1.5 (temperature/pressure compensated); ±1.7% (temperature compensated + external pressure compensation)
Mass flow (liquid): ±0.85% -
Measuring range
Liquid: 0.2 to 1895 m³/h (0.15 to 1115 ft³/min)
depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)
Steam, gas: 1.5 to 25285 m³/h (0.9 to 14880 ft³/min)
depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a) -
Max. process pressure
PN 250, Class 1500, 40K
-
Medium temperature range
Standard: –40 to +260 °C (–40 to +500 °F)
High/low temperature (option): –200 to +400 °C (–328 to +752 °F) -
Ambient temperature range
Compact version (standard): –40 to +80 °C (–40 to +176 °F)
Compact version (option): –50 to +80 °C (–58 to +176 °F)
Remote version (standard): –40 to +85 °C (–40 to +185 °F)
Remote version (option): –50 to +85 °C (–58 to +185 °F) -
Sensor housing material
Sensor connection housing: AlSi10Mg, coated; 1.4408 (CF3M)
-
Transmitter housing material
AlSi10Mg, coated; 1.4404 (316L)
-
Degree of protection
Compact version: IP66/67, type 4X enclosure
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, type 4X enclosure -
Display/Operation
4‐line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available -
Outputs
4‐20 mA HART (passive)
4‐20 mA (passive)
Pulse/frequency/switch output (passive) -
Inputs
4‐20 mA (passive)
-
Digital communication
HART, PROFIBUS PA, FOUNDATION Fieldbus, PROFINET over Ethernet-APL, Modbus TCP over Ethernet-APL
-
Power supply
DC 12 to 35 V (4‐20 mA HART with/without pulse/frequency/switch output)
DC 12 to 30 V (4‐20 mA HART, 4‐20 mA)
DC 12 to 35 V (4‐20 mA HART, pulse/frequency/switch output, 4‐20 mA input)
DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch output) -
Hazardous area approvals
ATEX, IECEx, cCSAus, JPN, EAC, UK Ex, KC
-
Product safety
CE, C-TICK, EAC
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation) -
Marine approvals and certificates
ABS, LR, BV, DNV, CCS
-
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
NACE MR0175/MR0103, PMI (on request); only Class 900/1500: welding test acc. to ISO 15614‐1, similar to ASME IX (on request)
-
-
Informações Técnicas (TI)
TI Proline Prowirl O 200
Versão em português - 11/2025
New version available in English
Medidor de vazão Vortex
EN PT 11/11/202506/05/202519/06/202421/06/202023/01/2020Idioma:
Português
Versão:
11/11/2025
Tamanho do arquivo:
4.9 MB
Nome do arquivo:
TI01334DPT_0725-00.pdf
-
-
-
Informações Técnicas (TI)
TI Proline Prowirl O 200
-
Informações Técnicas (TI)
Versão em português - 11/2025New version available in EnglishMedidor de vazão Vortex
EN PT 11/11/202506/05/202519/06/202421/06/202023/01/2020Idioma:
Português
Versão:
11/11/2025
Tamanho do arquivo:
4.9 MB
Nome do arquivo:
TI01334DPT_0725-00.pdf
-
-
-
Instruções de Operação (BA)
BA Proline Prowirl O 200 PROFIBUS PA
-
Instruções de Operação (BA)
Versão em português - 05/2025New version available in EnglishMedidor de vazão Vortex
EN PT 06/05/202501/01/2018Idioma:
Português
Versão:
06/05/2025
Tamanho do arquivo:
4.6 MB
Nome do arquivo:
BA01691DPT_0324-00.pdf
-
-
-
Instruções de Operação (BA)
BA Proline Prowirl O 200
-
Instruções de Operação (BA)
Versão em português - 11/2025New version available in EnglishMedidor de vazão Vortex
Modbus TCPEN PT 11/11/2025Idioma:
Português
Versão:
11/11/2025
Tamanho do arquivo:
3.5 MB
Nome do arquivo:
BA02399DPT_0125-00.pdf
-
-
-
Instruções de Operação (BA)
BA Proline Prowirl O 200 FOUNDATION Fieldbus
-
Instruções de Operação (BA)
Versão em português - 05/2025New version available in EnglishMedidor de vazão Vortex
EN PT 06/05/202501/01/2018Idioma:
Português
Versão:
06/05/2025
Tamanho do arquivo:
4.9 MB
Nome do arquivo:
BA01695DPT_0324-00.pdf
-
-
-
Instruções de Operação (BA)
BA Proline Prowirl O 200
-
Instruções de Operação (BA)
Versão em português - 05/2025New version available in EnglishMedidor de vazão Vortex
PROFINET em Ethernet-APLEN PT 06/05/202519/06/2024Idioma:
Português
Versão:
06/05/2025
Tamanho do arquivo:
4.5 MB
Nome do arquivo:
BA02134DPT_0225-00.pdf
-
-
-
Instruções de Operação (BA)
BA Proline Prowirl O 200 HART
-
Instruções de Operação (BA)
Versão em português - 05/2025New version available in EnglishMedidor de vazão Vortex
EN PT 06/05/202501/01/2018Idioma:
Português
Versão:
06/05/2025
Tamanho do arquivo:
4.5 MB
Nome do arquivo:
BA01687DPT_0324-00.pdf
-
-
-
Parâmetro do Equipamento (GP)
GP Proline Prowirl 200
-
Parâmetro do Equipamento (GP)
Versão em inglês - 11/2025New version available in EnglishVortex flowmeter
Modbus TCPEN PT 11/11/2025Idioma:
Inglês
Versão:
11/11/2025
Tamanho do arquivo:
1.7 MB
Nome do arquivo:
GP01240DEN_0125-00.pdf
-
-
-
Parâmetro do Equipamento (GP)
Proline Prowirl 200 HART Description des paramètres d'appareil
-
Parâmetro do Equipamento (GP)
Versão em francês - 01/2018New version available in EnglishDébitmètre vortex
EN PT 01/01/2018Idioma:
French
Versão:
01/01/2018
Tamanho do arquivo:
1.9 MB
Nome do arquivo:
GP01109DFR_0118.pdf
-
-
-
Instruções curtas (KA)
KA Medidor de vazão Proline 200
-
Instruções curtas (KA)
Versão em português - 06/2024New version available in EnglishTransmissor com sensor de vazão vortex
PROFINET com Ethernet-APLEN PT 19/06/2024Idioma:
Português
Versão:
19/06/2024
Tamanho do arquivo:
1.5 MB
Nome do arquivo:
KA01545DPT_0122-00.pdf
-
-
-
Instruções curtas (KA)
Flowmeter Proline 200 Brief Operating Instructions
-
Instruções curtas (KA)
Versão em inglês - 12/2018New version available in EnglishProfibus PA transmitter
with vortex flow sensorEN PT 03/12/201801/01/2018Idioma:
Inglês
Versão:
03/12/2018
Tamanho do arquivo:
1.7 MB
Nome do arquivo:
KA01328DEN_0218.pdf
-
Escolha uma das opções sugeridas
-
-- selected coding-- selected label99.999,00<%label%>
-
Procurando mais opções? Use o configurador individual para continuar.
-
-- selected coding-- selected label99.999,00<%feature%>
-
<%label%>99.999,00
-
Procurando mais opções? Use o configurador individual para continuar.
-
Process Connection
-
<%label%>99.999,00
-
<%label%>99.999,00
-
-
<%feature%><%coding%><%label%>
- Diagramas CAD Diagramas CAD
- Diagramas CAD
-
Preço por peça
-
<%range%><%price%>
-
- Acessórios opcionais
Configuração individual
Geralmente são fornecidos juntos
Acessórios
Produtos relacionados