Digital non-glass pH sensor
Tophit CPS491D

Tophit CPS491D - Digital non-glass pH electrode for heavily soiled media ©Endress+Hauser

The pH-value of a liquid can be calculated using the potentiometric measurement principle. This video shows what it is about and how this measuring principle works.

Memosens converts the measured value to a digital signal and transfers it to the transmitter without a contacting connection. This video shows how Memosens optimizes process control and maintenance.

Memosens ISFET sensor for heavily polluted media in chemical processes, paper or paint production

Tophit CPS491D is the specialist for fibrous media or media with a high content of suspended solids such as dispersions, precipitation, emulsions. Its unbreakable shaft ensures a long operating life while the open aperture guarantees reliable measurement under harsh process conditions. Thanks to Memosens digital technology, CPS491D combines maximum process and data integrity with simple operation. It resists corrosion and moisture, enables lab calibration and facilitates predictive maintenance.

  • Produtos simples

  • Fácil de escolher, instalar e operar

Excelência técnica

Simplicidade

  • Produtos padrão

  • Confiável, robusto e baixa manutenção

Excelência técnica

Simplicidade

  • Produtos com tecnologia de ponta

  • Altamente funcional e prático

Excelência técnica

Simplicidade

  • Produtos especializados

  • Desenvolvidos para aplicações exigentes

Excelência técnica

Simplicidade

Variável

Versões FLEX Excelência técnica Simplicidade
  • F
  • L
  • E
  • X

Versão Fundamental

Atende suas necessidades básicas de medição

Excelência técnica
Simplicidade
  • F
  • L
  • E
  • X

Versão Lean

Lida com seus principais processos facilmente

Excelência técnica
Simplicidade
  • F
  • L
  • E
  • X

Seleção Extended

Otimiza seus processos com tecnologias inovadoras

Excelência técnica
Simplicidade
  • F
  • L
  • E
  • X

Versão Xpert

Controla suas aplicações mais difíceis

Excelência técnica
Simplicidade

Variável

  • Especificações resumidas

    • Measurement range

      pH 0-14

    • Process temperature

      max. 130°C (266°F)

    • Process pressure

      max. 10 bar/100°C (145 psi/212°F)

Seleção rápida

Escolha uma das opções sugeridas
Preço de venda
  • <%label%>
  • <%feature%>
  • Process Connection
  • <%message%>

ou

Configuração individual

Continue sua seleção entre todas as opções possíveis
  • Benefícios

    • Unbreakable for highest product safety

    • Open aperture for application in heavily soiled media

    • Application possible at low temperatures

    • Low maintenance thanks to long calibration intervals

    • Maximum process safety through non-contact inductive signal transmission

    • Enables predictive maintenance due to storage of sensor and process-...

  • Campo de aplicação

    • Chemical processes

    • Pulp and paper industry

    • Contaminated media and suspended solids
      - Dispersions
      - Precipitation
      - Emulsions

    With ATEX, FM, and CSA approval for application in hazardous areas.

Características e especificações

Voltar

Details

  • Measuring principle

    ISFET

  • Application

    - Process applications - non-glass, break-proof - low temperatures - min. conductivity >500µS/cm - emulsions, suspensions, organic solvents

  • Characteristic

    - Rapid response at low temperatures - stabilised gel insensitive to contamination - lower amount of maintenance required than with glass - overhead installation

  • Measurement range

    pH 0-14

  • Measuring principle

    - Open aperture - double gel reference

  • Design

    - All shaft lengths with temperature sensor - advanced gel technology

  • Material

    PEEK / metal oxide / Perfluorelastomere

  • Dimension

    Diameter: 12 mm (0.47 inch)
    Shaft lengths: 120, 225, 360, 425 mm
    (4.72/8.86/14.2/16.7 inch)

  • Process temperature

    max. 130°C (266°F)

  • Process pressure

    max. 10 bar/100°C (145 psi/212°F)

  • Temperature sensor

    Pt1000.

  • Ex certification

    ATEX
    FM

  • Connection

    Contactless, inductive, digital connection head with Memosens technology

  • Ingres protection

    IP68

Documentos / Manuais / Software

Acessórios / peças sobressalentes

Nr
Order code
Specification
  • Successor