Tente buscar por palavras-chaves, código de pedido ou produto ou número de série, por exemplo, "CM442" ou "informações técnicas"
Insira pelo menos 2 caracteres para iniciar a busca.
  • Histórico

Memosens CLS16E

Seleção rápida

Escolha uma das opções sugeridas
Preço de venda
  • <%label%>
  • <%feature%>
  • Process Connection
  • <%message%>

ou

Configuração individual

Continue sua seleção entre todas as opções possíveis

Características e especificações

Voltar

Details

  • Measuring principle

    Conductive

  • Application

    Pure and ultrapure water

    Pharmaceutical industry
    Final rinse

    WFI

  • Characteristic

    Hygienic 2-electrode conductivity sensor

  • Measurement range

    k=0,1: 0,04 to 500 µS/cm

  • Measuring principle

    Conductive conductivity cell with electropolished stainless steel electrodes

  • Design

    Hygienic 2-electrode conductivity sensor with coaxially arranged electrodes, electropolished

  • Material

    Isolation: PEEK
    Electrode: polished stainless steel 1.4435
    Sealing: Isolast (FFKM) FDA approved

  • Dimension

    Electrode diameter: 17 mm (0.66 inch)
    Electrode length: 54 mm (2.13 inch)

  • Process temperature

    -5 to 120 °C (23 to 248 °F)
    For Sterilizsation: max. 150 °C at 5 bar
    (Max. 302 °F at 73 psi)

  • Process pressure

    13 bar at 20 °C (188 psi at 68 °F)
    9 bar at 120 °C (130 psi at 248 °F)

  • Temperature sensor

    Pt1000

  • Ex certification

    ATEX, NEPSI, CSA, IECEx, INMETRO, EAC Ex

  • Connection

    Process connection: Varivent, Neumo, BioControl, Clamp
    Sensor connection: Inductive, digital connection head with Memosens 2.0 technology

  • Ingres protection

    IP68

  • Additional certifications

    Material certification 3.1
    EHEDG certified, hygienic design
    ASME BPE-2002
    Conformity to cGMP derived

Documentos / Manuais / Software

Acessórios / peças sobressalentes

Nr
Order code
Specification
    • Order code 50081902

      Calibration solution CLY11-A, 74,0 uS/cm

      500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO

    • Order code 50081903

      Calibration solution CLY11-B,149,6 uS/cm

      500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO.

    • Order code 50081904

      Calibration solution CLY11-C,1,406mS/cm

      500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO

    • Order code 50081905

      Calibration solution CLY11-D,12,64 mS/cm

      500 ml, reference temperature 25oC. Precision solution referred to SRM to NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO

    • Order code 50081906

      Calibration solution CLY11-E,107,00mS/cm

      500 ml, reference temperature 25oC. Precision solution referred to SRM of NIST. With temp. table for qualified calibration of conductivity measuring systems acc. to ISO